7.1

Temperature Coefficient sub-function

This sub-function allows you to select correct temperature coefficient for optimum operations. For applications
in pure water or ultra-pure water industries, select “Pur” temperature coefficient option. For all other applications,
select “Lin” temperature coefficient. Controller allows further input of temperature coefficient values,
independently for process and for calibrating solutions. Default is “Pur”.

HOLD SETUP

HOLD SETUP

HOLD SETUP

HOLD SETUP

tc

Lin

tC

HOLD SETUP

2.10

P.tC

2.10

C.tC


%

ATC


%

ATC

Pur

tC


7.1.1

Selecting Pure-water or Linear Temperature Coefficient

tc

HOLD SETUP

1) Enter Advanced set-up mode. Push ENTER key and scroll to

Advanced Set-up security code “22”. Push ENTER key. Controller
displays “tC”. Press ENTER key again. Controller displays “Pur” in upper
display and “tC” in lower display. Default is “Pur”.

2) Press

Δ or keys to select between “Pur” and “Lin” temperature

coefficients. If “Lin” tC is to be selected, press ENTER key when
controller displays “Lin”.

Lin

tC

HOLD SETUP

3) If “Lin” is selected, upper display shows “2.10%”, while lower display

show “P.tC”. This option allows input of Process Temperature coefficient,
from 0 to 10%. Default is 2.10%. Use

Δ or keys to enter required

temperature coefficient value.

4) Press ENTER key. Next, enter the calibration solution temperature

coefficient value, “C.tC”, from 0 to 10%. Default is 2.10%.

2.10

%

P.tC

ATC

HOLD SETUP

5) Press

Δ or keys to input desired calibration solution temperature

coefficient.

6) Press ENTER key to accept.
7) Continue with additional Advance setup procedures or return to

Measurement mode by pressing

Δ and keys (escape) simultaneously.

HOLD SETUP

2.10

%

C.tC

ATC