Alpha RES 1000

16

1) Follow directions in 7.4.1 to enter Controller mode.
2) Press the ENTER key. The upper display shows the current controller type and the lower display shows

“tYP”.

3) Press the

Δ and keys to select your controller type.

-

L.Ct = limit value pickup (on/off control).

-

oFF = controller off.

-

PLC = pulse length control.

-

PFC = pulse frequency control.

4) Press the ENTER key to confirm your selection.
5) Proceed to 7.4.3 step 3, or return to Measurement mode by pressing the

Δ and keys simultaneously

(escape).


7.4.3

Choosing break/make contact relay type

Note: If the controller type “oFF” is set, the parameters listed in 7.4.4 and 7.4.5 are blanked out.
This mode lets you determine the relay-state under Non-Alarm condition – dEEN (de-energised) or EN
(energised).

1) Follow directions in 7.4.1 to enter Controller mode.

dEEn

rEL

HOLD SETUP

2) Press the ENTER key. Scroll until the lower display shows “rEL” and the

upper display shows the current selection (de-energised = dEEN or
energised = EN).

3) Press the

Δ or keys to choose de-energised or energised relay state.

4) Press the ENTER key to confirm your selection.
5) Continue with Advanced Set-up mode procedures, or return to

Measurement mode by pressing the

Δ and keys simultaneously

(escape).

En

rEL

HOLD SETUP



7.4.4

Selecting proportional range value Xp

Note: If the controller type “oFF” or “L.Ct” is set, the parameters listed in 7.4.4 and 7.4.5 are blanked out.
This mode lets you set a band as a percentage of its full scale value. You can select this range from 10 to
200%, and the lower display shows “PrP”.

1) Follow directions in 7.4.1 to enter Controller mode.

100

PrP

HOLD SETUP

2) Press the ENTER key. Scroll until the upper display shows the

proportional range (a number from 10 to 200%), and the lower display
shows “PrP”.

3) Press the

Δ and keys to choose the proportional range value Xp.

4) Press the ENTER key to confirm your selection.

5) Proceed to 7.4.5 step 3, or return to Measurement mode by pressing the

Δ and keys simultaneously

(escape).