6 viewing electrode properties – Thermo Fisher Scientific Alpha pH 550 User Manual

Page 34

Advertising
background image

- 30 -

CAL dEF:

Reset all calibration settings of the pH Monitor to
factory defaults, when confirmed by pressing ENT
key.

Press ENT key to confirm your selection.

- If ‘nO dEF’ was selected, the pH Monitor reverts to Cnfg screen.

Press

▲ or ▼ key to access other setup screens or

press

▲ and ▼ key simultaneously (escape) to return

to measurement mode.

- If ‘FCt dEF’ or ‘CAL dEF’ was selected, the pH Monitor performs
the selected reset and returns to measurement mode.

Press

▲ or ▼ key to access other setup screens or press ▲ and ▼ key

simultaneously (escape) to return to measurement mode.

To exit from any intermediate steps, press

▲ and ▼ keys simultaneously

(escape). The pH Monitor returns to the first screen configuration settings (Cnfg).

5.6

Viewing Electrode Properties

Each time you calibrate your pH electrode, the pH Monitor re-calculates slope &
offset of the electrode and shown in the LCD at the end of calibration. The setup
mode allows you to view slope & offset values at any time.

NOTE:

When the measurement mode is set to ORP, only the offset value of
the electrode is displayed.

SETUP

HOLD

mV

pH

2

SETUP

ENT

1

ENT

mV

2

SETUP

For pH measurement mode

For ORP measurement mode

ENT