Thermo Fisher Scientific Alpha pH 550 User Manual

Page 36

Advertising
background image

- 32 -

6

T E C H N I CAL S P E CI F I C AT I O NS

General Specification

(a) pH

Measuring Range

0.00 to 14.00 pH

Resolution 0.01

pH

Accuracy

± 0.01 pH

(b) mV

Measuring Range

-1000 to 1000 mV

Resolution 1

mV

Accuracy

± 1 mV

(c) Temperature

Measuring range

-10.0 to +125.0

o

C or +14.0 to +257

o

F

Resolution 0.1

o

C

0.1

o

F (Resolution is 0.1

o

F up to 199.9

o

F and

1

o

F for 200

o

F and above)

Relative accuracy

±0.5

o

C

±1.0

o

F

Sensor

2 wire / 3 wire - Pt100 or Pt1000 (Jumper
Selectable)

Compensation

Automatic or Manual

Calibration

(a) pH

Number of calibration points

1 or 2 points

Number of calibration buffers

USA: 1.68,4.01, 7.00, 10.01,12.45
NIST: 1.68,4.01, 6.86, 9.18 ,12.45

(b) Temperature

Offset Adjustment

± 10 ˚C/ ± 18 ˚F

Input /Output

Input Asymmetrical

/

Symmetrical

Output None

Display

LCD

7 segments display with symbols for status
information

Electrical Data and Connections

pH/mV Input

Screw Terminal (3.5mm pitch)

Connection terminal

8-pin & 2-pin terminal blocks

Other

Power Input

Universal +9 V DC Adaptor and 24V DC
Supply

Dimensions (W x H x D)

96mm x 96mm x 66 mm

Weight (Estimated)

210g

Ambient Temp. operating range

0 to 40

o

C

Maximum Relative Humidity

80% up to 31

o

C decreasing linearly to 50% at

40

o

C

Advertising