8 T ROUBLE SHOOT IN G

Problem

Cause

Solution

Power on, but no display

a)

Loose connections

b)

Incorrect output voltage of
the power adaptor

a)

Ensure cables make good
contact

b)

Use an power adaptor with
specified output voltage

Unstable pH reading

a)

Dirty electrode

b) Electrical

noise

interference

a)

Clean electrode and
recalibrate

b)

Switch to Symmetric mode

Oscillating temperature
readings

a) Electrical

noise

interference

a)

Ensure shield wire is
properly connected to pin 7

Slow response

a) Dirty / Oily electrode

a) Clean electrode

Blinking ATC

a)

No temperature probe
connection during ATC
mode

a)

Ensure temperature sensing
cable makes good contact

Blinking electrode
annunciator

a)

Error in calibration

a)

Ensure calibration standard
solution is not contaminated.
Ensure electrode is clean.

Or (pH)

a)

pH electrode is not
connected

a)

Ensure electrode makes
good contact with
Transmitter

Or (Temperature)

a)

Temperature probe is not
connected when ATC
enabled

a)

Ensure electrode makes
good contact with
Transmitter