5. Calibrating the Controller

5.1 The Lower Level Menus

The "CAL PH/CAL OrP" display is the first menu seen upon
entering the lower-level menus, depending on which mode is
set. Press ▲ or ▼ key to display the various menus as shown
in the figure below. Press ENTER key to go to the lower level
menus.

5.2 Calibrating for pH Measurement

5.2.1 The CAL PH menu

1. Enter this menu by pressing the MODE key twice to the

“SEt” display if you are in measurement mode.

2. Press

ENTER key. If the password has been set earlier,

key in the password using the method described in section
4.1. Press ENTER once to confirm. After you set the
correct password, you can see the “CAL” and “PH
displays, blink alternately.

3. Press

ENTER key to enter into the pH calibration mode.

Dip both the pH electrode and temperature probe in the
standard solution for a few minutes.

4. When

CAL PH” displays, press ENTER key. The current

measured value (7

±

0.3 pH) will be displayed. Once the

reading stablizes, press ENTER to confirm.

5. Use

▼ keys to scroll to the other buffer values “PH4” and

PH10”, in that sequence. Press ENTER to confirm and

START

CAL pH /
CAL OrP

CAL