5.4 Calibrating for ORP Measurement

1.

Make sure that the ORP mode is
set before calibration.

2.

Enter this menu by pressing
MODE key twice to the “SEt”
display if you are in measurement
mode.

3. Press

ENTER key. If the password

has been set earlier, key in the
password using the method
described in section 4.1. Press
ENTER once to confirm. After you
set the correct password, you can
see the “CAL” and “OrP” displays
blink alternately.

4. Press

ENTER key to enter into the

ORP calibration mode. Dip both
the ORP electrode and temperature probe in the standard
solution for a few minutes.

5. When

CAL OrP” displays, you can see the present mV

value blinking, press ▲ or ▼ key to adjust the mV value.
To confirm press ENTER key.

6. Press

MODE key to return the controller to its

measurement mode.