5.6.2 The Hi HYS Menu

1.

Enter this menu by pressing the
MODE key to the “SEt” display
if you are in the measurement
mode.

2. Press

ENTER key. If the

password has been set earlier,
key in the password using the
method described in section
4.1. Press ENTER once to
confirm. After you set the
correct password, you see the
CAL” and “PH” display blinks
alternatively. Press ▼ key
thrice, the display shows “Hi”
“HYS”
blinking.

1. Press

ENTER key to access

the Hi HYS menu and the
display shows the last Hi HYS

value. Use ▲ and ▼ keys to
change the value of the Hi HYS
point.

2. Press ENTER to confirm the

value of the Hi HYS.

3.

Press MODE key to exit to
measurement mode. You can
press MODE key (as an
ESCAPE key) to revert to the
measurement value if ENTER is
not pressed; the set value is not
stored into memory.

SETUP

MODE

ENTER

Hi

SETUP

MODE

ENTER

0.2

SETUP

MODE

ENTER

HYS

NOTE : The
maximum value of
Hysteresis is 0.2 pH
to 2.0 pH or 20 mV to
200 mV. The HI HYS
hysteresis is spread
equally on either side
of the Hi SET point.