INSTALLATION

Prior to installation, the flow meter should be checked internally for foreign material and to ensure the turbine rotor spins

freely . Fluid lines should also be checked and cleared of all debris .
The flow meter must be installed with the flow arrow, etched on the exterior of the meter body, pointing in the direction of

fluid flow . Though the meter is designed to function in any position, it is recommended, where possible, to install horizontally

with the conduit adapter facing upward . Thread a magnetic pickup (Blancett model B111109 or equivalent) into the conduit

adapter completely finger tight without forcing . Secure with a lock nut if supplied . Install conduit or other fittings suitable for

the installation area onto the conduit adapter hub on the flow meter .
The liquid being measured should be free of any large particles that may obstruct rotation of the rotor . If particles are present,

a mesh strainer should be installed upstream before operation of the flow meter .

Part Number

Strainer Mesh

Clearance

Filter Size

B131C-038

60 × 60

0 .0092

260 Micron

B131C-050

60 × 60

0 .0092

260 Micron

B131C-075

60 × 60

0 .0092

260 Micron

B131C-088

60 × 60

0 .0092

260 Micron

B131C-100

60 × 60

0 .0092

260 Micron

B132C-050

60 × 60

0 .0092

260 Micron

B132C-075

60 × 60

0 .0092

260 Micron

B132C-088

60 × 60

0 .0092

260 Micron

B132C-100

40 × 40

0 .0150

400 Micron

B132C-150

20 × 20

0 .0340

0 .86 mm

B132C-200

20 × 20

0 .0650

0 .86 mm

B133C-300

10 × 10

0 .0900

1 .6 mm

B134C-400

10 × 10

0 .0650

1 .6 mm

B136C-600

4 × 4

0 .1875

4 .8 mm

B138C-800

4 × 4

0 .0900

4 .8 mm

B139C-900

4 × 4

0 .0900

4 .8 mm

Table 1: Strainer mesh installation details

The preferred plumbing setup is one containing a bypass line (

Figure 4

) that allows meter inspection and repair without

interrupting flow . If a bypass line is not used, it is important that all control valves be located downstream of the flow meter

(

Figure 5

) .

PRESSURE IN EXCESS OF ALLOWABLE RATING MAY CAUSE THE HOUSING TO BURST AND CAUSE SERIOUS PERSONAL

INJURY.

DAMAGE CAN BE CAUSED BY STRIKING AN EMPTY METER WITH A HIGH VELOCITY FLOW STREAM.

Turbine Flow Meter, Explosion-Proof QuikSert