Calibration, Calibration tables – Badger Meter Irrigation Sensors - 1/2 and larger" User Manual

Page 8

Advertising
background image

CALIBRATION

Badger Meter sensors use unique K and Offset numbers for calibration. These numbers are derived from calibration runs using
NIST traceable instruments. Using both a K and an Offset number provides higher accuracy than using a K (pulse/gal) factor
alone. K and Offset numbers for each tee configuration are listed in the following tables.

Calibration Tables

The table below provides calibration and operation data for Data Industrial Metal Tee Sensors 0.5 inch to 2.5 inch.

Column 1

Tee Model Number

Column 2

Apparent I.D. - For use with the Series 900 and the Series 1000

Columns 3 and 4

The K value and Offset values to use in our frequency equation

This equation describes the frequency of the output signal of all Badger Meter flow sensors. By
substituting the appropriate K and Offset values from the table, the sensor’s output frequency can
be calculated for each pipe size. This information is required when calibrating an output board
or when using the raw sensor data as direct output to interface with a device that is not a Badger
Meter product.

Column 5

This column indicates the suggested flow range of each tee sensor. Badger Meter sensors will
operate both above and below the indicated flow rates. However, good design practice dictates the
use of this range for best performance.

Sensors should be sized for flow rather than pipe size. To prevent disturbances to the flow profile
always connect the sensor tee to pipe nipples measuring at least 10 pipe diameters in length on the
up stream (supply) side and at least 5 pipe diameters in length on the downstream (delivery) side
before making the transition in pipe size.

Calibration Table for

Series 228BR, 228CB, 250BR, 228SS

Apparent I.D.

Suggested

Model for

Series

K

Offset

Operating

1000, 900

Value

Range
(GPM)

228BR20xx-xxxx 1.99 2.747 0

10-100

228BR25xx-xxxx 2.52 3.741

0.386 16-160

228CB20xx-xxxx 2.07 2.809

0.276 12-120

150 PSI Tee

2.07

400 PSI Tee

2.1

2.604

0.25

12-120

228CB25xx-xxxx 2.51 3.74

0.277 16-160

228SS20xx-xxxx

1.99

2.747

0

10-100

250BR05xx-xxxx

sch40 PVC

none

0.337379

0.097041

0.8-8

sch80 PVC

none

0.338073

0.134854

0.8-8

sch40 steel

none

0.356212

0.075729

0.8-8

Type L

none

0.350899

-0.321666

0.8-8

250BR07xx-xxxx

sch40 PVC

none

0.436827

0.567915

1-10

sch80 PVC

none

0.43983

0.692372

1-10

sch40 steel

none

0.434836

0.766196

1-10

Type L

none

0.432127

0.619813

1-10

250BR10xx-xxxx

1.05

0.397368

0.261768

2-40

250BR12xx-xxxx

1.38

0.76447

0.16489

3-60

250BR15xx-xxxx

1.61

1.06526

0.0892

4-80

250 and 228 Metal Tee Impeller Flow Sensors

Page 8

August 2012