PVC schedule 80 pipe of the appropriated diameter is installed upstream of the sensor, and a similar 5 to 10 diameter lengths
downstream.

PVDF sensor calibration is based on flow through industry standard PVDF pipe. Similar action should be taken if the PVDF
sensor is used in a system based on other piping technologies.

CALIBRATION TABLE-DIGITAL OUTPUT

The following table provides calibration and operational data for the various models of the Badger Series 4000 sensor. The
data is organized as follows:

Column 1:

Model number, to which the data listed in the particular column pertains

Column 2:

Housing material

Column 3:

Nominal pipe size

Column 4:

Pipe O.D., per applicable standards

Column 5:

Pipe I.D., per applicable standards

Column 6:

Slope of the regression line used in calibration (K)

Column 7:

Intercept of the regression line (Offset)

Column 8:

Minimum recommended flow rate (GPM)

Column 9:

Maximum recommended flow rate (GPM)

Column 10:

Pulse rate at minimum flow (Hz)

Column 11:

Pulse rate at maximum flow (Hz)

The calibration constants in columns 6 and 7 relate frequency (Hz) to flow rate (GPM) in the equations:

FREQ= GPM

K -Offset

GPM = K (FREQ + Offset)