CALIBRATION

Badger Meter Impeller sensors use unique K and offset numbers for calibration. These numbers are derived from calibration

runs using NIST traceable instruments. Using both a K and an offset number provides higher accuracy than using a K factor

alone. The K and offset numbers for each tee configuration are listed in the "Calibration Table" below.

Calibration Table Columns

The table below provides calibration and operation data for Badger Meter Plastic Tee Sensors 1-1/2" to 4".

Column 1

Sensor Model Number

Columns 2 and 3 The K value and Offset values to use in our frequency equation:

Freq= Gpm

K

- offset

This equation describes the frequency of the output signal of all Badger Meter flow sensors. By

substituting the appropriate K and Offset values from the table, the sensor’s output frequency can be

calculated for each pipe size. This information is required when calibrating an output board or when

using the raw sensor data as direct output to interface with a device that is not a Badger Meter product.

Calibration Table

Size/Schedule

K

Offset

1/2" S40

0.078000

0.9

1/2" SDR 13.5

0.120119

0.1

3/4" S40

0.156300

0.9

3/4" SDR 21

0.197000

–0.6

1" S40

0.261119

1.2

1" SDR 21

0.321739

0.6

Series 735PV Plastic Tee Type Impeller Flow Sensor