Communications cable wiring

Field calibration requires a Data Industrial A320 Pro-

gramming kit (consisting of a custom cable and soft-

ware) and a PC running Windows® 9x, ME, NT, 2000 or

XP. In order to program, the

Model 320 must be connect-

ed to power, and the A301

cable must be connected to

the Model 320 Comm port

connector and an available

DB9 COM port on a com-

puter.

Note:

The Data Industrial A301 Cable will work with all 300

Series products. However the older version of the cable

(A300) does not have sufficient bandwidth to work with

the newer 340 Series Transmitters or SDI Flow Sen-

sors.

Programming Software Installation

Floppy Installation

Place the software installation Disk 1 into the floppy

drive and run the setup.exe program to install.

CDROM Installation

Software CD into the CDROM drive and it should auto-

start then Click software and click the product name for

what software you want installed and the installation will

begin.

Web Installation

Data Industrial provides free programming software up-

dates via the Internet for all of 300 Series devices. The

Installation software can be found at the Data Industrial

web site (www.dataindustrial.com) in the support sec-

tion.

Model 320 Programming

Programming the Model 320 is accomplished by install-

ing the Data Industrial programming software on a com-

puter and entering data on templates of the Windows®

based program.

1. Load the interface software into the computer.

2. Connect the computer to the Model 320 transmit-

ter using the Data Industrial A301 communications

cable. Plug A301 cable to the socket labeled “D.I.C

Comm Port” taking care to properly align the tab on the

plug and socket to maintain polarity then plug the DB9

connector of the Data Industrial A301 communications

cable to an avaliable PC com port that has the Model

320 software installed.

3. Connect the Model 320 transmitter to a power sup-

ply.

Data Industrial

Model 320 Installation Guide

3

Communications cable wiring

Field calibration requires a Data Industrial A320 Programming kit

(consisting of a custom cable and software) and a PC running Windows

®

9x, ME, NT, 2000 or XP. In order to program, the Model 320 must be

connected to power, and the A301 cable must be connected to the Model

320 Comm port connector and an available DB9 COM port on a

computer.
Note:

The Data Industrial A301 Cable will work with all 300 Series

products. However the older version of the cable (A300) does not have sufficient bandwidth to

work with the newer 340 Series Transmitters or SDI Flow Sensors.

Programming Software Installation
Floppy Installation

Place the software installation Disk 1 into the floppy drive and run the setup.exe program to install.

CDROM Installation

Software CD into the CDROM drive and it should autostart then Click software and click the product

name for what software you want installed and the installation will begin.

Web Installation

Data Industrial provides free programming software updates via the Internet for all of 300 Series

devices. The Installation software can be found at the Data Industrial web site (www.dataindustrial.com)

in the support section.

Model 320 Programming
Programming the Model 320 is accomplished by installing the Data Industrial programming software on

a computer and entering data on templates of the Windows

®

based program.

1. Load the interface software into the computer.
2. Connect the computer to the Model 320 transmitter using the Data Industrial A301 communications

cable. Plug A301 cable to the socket labeled “D.I.C Comm Port” taking care to properly align the tab

on the plug and socket to maintain polarity then plug the DB9 connector of the Data Industrial A301

communications cable to an avaliable PC com port that has the Model 320 software installed.

3. Connect the Model 320 transmitter to a power supply.
4. Open the interface software and select the appropriate COM PORT as shown in the dialog box

below.

Location of the

DIC Communication Port

S

P

ow

er

In

C +

C -

D.I.C.

Comm

Port

Input led

DIC

Communications

Port

Location of the

DIC Communication Port

Data Industrial

Model 320 Installation Guide

3

Communications cable wiring

Field calibration requires a Data Industrial A320 Programming kit

(consisting of a custom cable and software) and a PC running Windows

®

9x, ME, NT, 2000 or XP. In order to program, the Model 320 must be

connected to power, and the A301 cable must be connected to the Model

320 Comm port connector and an available DB9 COM port on a

computer.
Note:

The Data Industrial A301 Cable will work with all 300 Series

products. However the older version of the cable (A300) does not have sufficient bandwidth to

work with the newer 340 Series Transmitters or SDI Flow Sensors.

Programming Software Installation
Floppy Installation

Place the software installation Disk 1 into the floppy drive and run the setup.exe program to install.

CDROM Installation

Software CD into the CDROM drive and it should autostart then Click software and click the product

name for what software you want installed and the installation will begin.

Web Installation

Data Industrial provides free programming software updates via the Internet for all of 300 Series

devices. The Installation software can be found at the Data Industrial web site (www.dataindustrial.com)

in the support section.

Model 320 Programming
Programming the Model 320 is accomplished by installing the Data Industrial programming software on

a computer and entering data on templates of the Windows

®

based program.

1. Load the interface software into the computer.
2. Connect the computer to the Model 320 transmitter using the Data Industrial A301 communications

cable. Plug A301 cable to the socket labeled “D.I.C Comm Port” taking care to properly align the tab

on the plug and socket to maintain polarity then plug the DB9 connector of the Data Industrial A301

communications cable to an avaliable PC com port that has the Model 320 software installed.

3. Connect the Model 320 transmitter to a power supply.
4. Open the interface software and select the appropriate COM PORT as shown in the dialog box

below.

Location of the

DIC Communication Port

S

P

ow

er

In

C +

C -

D.I.C.

Comm

Port

Input led

DIC

Communications

Port

4. Open the interface software and select the appro-

priate COM PORT as shown in the dialog box below.

5. Open the Parameters Screen as shown below.

6. Program using diagram below as a reference

Note #1

Sdi - If the SDI sensor type is selected the required

K and offset values can be found the the SDI owners

manual.

4000 - If the 4000 sensor type is selected, click the

choose button and select the sensor from the pull down

box that appears.

Sine - Provided for connection to sensors which have a

sine wave output. Please consult sensor manufacturer

for the calibration settings.

200 Insert Type - If the 200 Insert Sensor type is

selected the required K and offset can be found the

the 200 owners manual or if the manual is not handy

the calculate button can be pushed and an inside pipe

diameter can be entered and once calculate is pressed

a K and offset will automatically be entered in.

200 Tee Type - If the 200 tee type is selected, click the

choose button and select the sensor from the pull down

box that appears.

Data Industrial

Model 320 Installation Guide

4

6. Program using diagram below as a reference.

Note #1

Sdi - If the SDI sensor type is selected the required K and offset values can be found the the SDI

owners manual.
4000 - If the 4000 sensor type is selected, click the choose button and select the sensor from the pull

down box that appears.
Sine - Provided for connection to sensors which have a sine wave output. Please consult sensor

manufacturer for the calibration settings.
200 Insert Type - If the 200 Insert Sensor type is selected the required K and offset can be found the

the 200 owners manual or if the manual is not handy the calculate button can be pushed and an inside

pipe diameter can be entered and once calculate is pressed a K and offset will automatically be entered

in.
200 Tee Type - If the 200 tee type is selected, click the choose button and select the sensor from the

pull down box that appears.

To go to calibration settings

screen select "Parameters"

from either place shown

5. Open the Parameters Screen as shown below.

Step #2

Select Flow Rate Units.

Step #1

See Note #1

Select Sensor Type and

Enter K and offset numbers.

Step #3

Enter units per pulse and select

pulse width.
Step #4

Send

Press

to transmit

calibration data to the 320.

Press to refresh the parameters

screen with the current 320 settings.

Press to restore the factory defaults

To save the factory defaults Send

must be pressed before values take effect.

Step #5

Press to exit Parameters screen and

go back to the main screen.

Data Industrial

Model 320 Installation Guide

4

6. Program using diagram below as a reference.

Note #1

Sdi - If the SDI sensor type is selected the required K and offset values can be found the the SDI

owners manual.
4000 - If the 4000 sensor type is selected, click the choose button and select the sensor from the pull

down box that appears.
Sine - Provided for connection to sensors which have a sine wave output. Please consult sensor

manufacturer for the calibration settings.
200 Insert Type - If the 200 Insert Sensor type is selected the required K and offset can be found the

the 200 owners manual or if the manual is not handy the calculate button can be pushed and an inside

pipe diameter can be entered and once calculate is pressed a K and offset will automatically be entered

in.
200 Tee Type - If the 200 tee type is selected, click the choose button and select the sensor from the

pull down box that appears.

To go to calibration settings

screen select "Parameters"

from either place shown

5. Open the Parameters Screen as shown below.

Step #2

Select Flow Rate Units.

Step #1

See Note #1

Select Sensor Type and

Enter K and offset numbers.

Step #3

Enter units per pulse and select

pulse width.
Step #4

Send

Press

to transmit

calibration data to the 320.

Press to refresh the parameters

screen with the current 320 settings.

Press to restore the factory defaults

To save the factory defaults Send

must be pressed before values take effect.

Step #5

Press to exit Parameters screen and

go back to the main screen.