Badger Meter 340N2 Series Metasys Btu Transmitter User Manual

Page 6

Advertising
background image

6

Step 1

see note #1

Select the flow sensor type

(sine or pulse) and enter the

K and Offset -

Step 4

Select the desired flow

rate and total units here

Step 7

Press send to transmit

calibration data to the
Series 340N2

Step 5

Select the desired energy

rate and total units here

Step 2

Select the desired temperature

sensor units.

Step 3

See Note #2

Typically:

Select the method of computing the

temperature differential

T1>T2 for Heating
T1<T2 for Cooling

Step 6

Select the output units per

pulse, and the pulse width

Press to restore the factory default

settings to the screen

must press “send” before values

take effect

Note:

Step 8

Press to exit parameters screen

and to go back to main screen

Press to refresh the parameters screen with

current Series 340N2 settings

Will overwrite any changes not sent to Series 340N2.

See Note #3

Step 1

Select a new device address using the

New Device Address pull down box.

Step 2

Press the Set New Device Address

button to send a new N2 address to

the Series 340N2.

Step 3

Press the Exit button to return

to the Parameters screen

Note #1:
Badger Meter sensors are pulse type sensors. The K and
Offset information is printed in the owners manual shipped with
the product. This information is also available on our website.
Calibration constants for other sensors must be supplied by the
manufacturer.

Note #2
Typically the temperature measured by T1 will be greater
than T2 in a heating application and less than T2 in a cooling
application. The selection of one of these choices will determine

if energy calculations are made for heating only (T1>T2),

cooling only (T1<T2), or both (absolute).

Note #3

The filter coefficient screen allows adjustment of the flow and

energy filters. A scale of 0-10 is used with 10 providing the

greatest degree of smoothing. Settings greater then “10” are
permitted by selecting other. However, settings greater then

“10” should be used with extreme caution as it will cause the

Series 340N2 to respond very slowly to actual changes in flow

and energy. This could cause misreporting of fast changing flow
and energy consumption data. See the dialog box below.

Advertising