LOCATION INSTRUCTIONS

Straight pipe requirements: Accuracy is affected by the piping configurations due to the disturbances of the flow profile . A

fully developed symmetrical flow profile is achieved with the minimum upstream and downstream recommended lengths .

DISTANCES ARE EXPRESSED IN

NUMBER OF NOMINAL PIPE DIAMETERS

THROTTLED VALVE OR REGULATOR

TWO ELBOWS IN DIFFERENT PLANES

CHANGE IN PIPE SIZE

TWO ELBOWS IN THE SAME PLANE

A SINGLE ELBOW

Figure 11: Location instructions