Sync mode, 4 sync mode – PNI TCM User Manual

Page 37

Advertising
background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 32

6.5.4

Sync Mode

Sync Mode enables the module to stay in Sleep Mode until the user’s system sends a

trigger to report data. When so triggered, the TCM will wake up, report data once, then

return to Sleep Mode. One application of this is to lower power consumption. Another

use of the Sync Mode is to trigger a reading during an interval when local magnetic

sources are well understood. For instance, if a system has considerable magnetic noise

due to nearby motors, the Synch Mode can be used to take measurements when the

motors are turned off.

Enter Sync Mode

On the Test tab, above the tabs and 3D model, click the “Sync Mode” check box to

enter Sync Mode.

Sync Mode Output

To retrieve the first reading, click the <Sync Read> button. Heading, pitch and roll

information will be displayed on Current Reading window. If the “Enable 3D Model
Display” box is selected on the Configuration tab, then the helicopter will follow the
movement as well. The module will enter Sleep Mode after outputting the heading,

pitch, and roll information. To obtain subsequent readings, the user should first click

on the <Sync Trigger> button to wake up the module and then click on the <Sync

Read> button to get the readings, after which the module will return to sleep.

Exit Sync Mode

Click on the <Sync Trigger> button and then uncheck the “Sync Mode” check box to

exit Sync Mode.

Note that <Sync Trigger> sends a 0xFF signal as an external interrupt to wake up the

module. This is not done for the first reading as the module is already awake.

Advertising