Ksetconfigdone (frame id 19, Ksetfirfiltersdone (frame id 20, Kstartcontinuousmode (frame id 21 – PNI TCM User Manual

Page 59: Kstopcontinuousmode (frame id 22, 19 ksetconfigdone (frame id 19, 20 ksetfirfiltersdone (frame id 20, 21 kstartcontinuousmode (frame id 21, 22 kstopcontinuousmode (frame id 22

Advertising
background image

PNI Sensor Corporation

DOC#1014688 r06

TCM User Manual

– July 2013

Page 54

DistError:

For a magnetic calibration, this score indicates if the distribution of sample points is

sufficient, with an emphasis on the heading distribution. The score should be 0.

Significant clumping or a lack of sample points in a particular section can result in a

poor score. In the event of an aborted calibration the score will be 179.8

d

, or in the

event of an accel-only calibration the score will be 99.99

d

.

TiltError:

This score indicates if the TCM experienced sufficient tilt during a magnetic

calibration, taking into account the calibration method. The score should be 0. In the

event of an aborted calibration the score will be 179.8

d

, or in the event of an accel-

only calibration the score will be 99.99

d

.

TiltRange:

For a magnetic calibration, this reports the larger of either half the full-pitch range or

half the full-roll range of sample points. For example, if the device is pitched +10° to

-20º, and rolled +25º to -15º, the TiltRange value would be 20º, which represents half

the roll range. For Full-Range Calibration and Hard-Iron-Only Calibration, this
should be ≥45°. For 2D Calibration, ideally this should be ~2°. For Limited Tilt
Range Calibration the value should be as large a possible given the user’s constraints.
In the event of an aborted calibration the score will be 179.8

d

, or in the event of an

accel-only calibration the score will be 99.99

d

.

7.3.19 kSetConfigDone (frame ID 19

d

)

This frame is the response to kSetConfig frame. The frame has no payload.

7.3.20 kSetFIRFiltersDone (frame ID 20

d

)

This frame is the response to kSetFIRFilters. The frame has no payload.

7.3.21 kStartContinuousMode (frame ID 21

d

)

If the TCM is configured to operate in Continuous Acquisition Mode, as defined by

kSetAcqParams, then this frame initiates the outputting of data at a relatively fixed data

rate, where the data rate is established by the SampleDelay parameter. The frame has no

payload. The response is kGetDataResp.

7.3.22 kStopContinuousMode (frame ID 22

d

)

This frame commands the TCM to stop data output when in Continuous Acquisition

Mode. The frame has no payload.