a. Note the warning and press RETURN

b. Type “C” and press RETURN to copy the current HygroGen2 settings to your USB installation

drive and to confirm the action type “Y” and press RETURN.

This creates a snapshot of the current HygroGen2 settings, particularly customisations to the

controller settings that may have been tuned uniquely for your particular instrument. At the end

of the installation, the USB stick should be labelled with the serial number of the HygroGen2 and

stored safely. This USB stick should be used in the event the unit requires a system recovery or

settings restore.

c. Type Y to install HygroGen2 version 2.0 software.

d. Confirm no other USB peripherals apart from keyboard, mouse and your USB installation

drive are plugged in. Press any key to continue.

e. The installation process will then proceed and will take approximately 30 minutes.

Select appropriate settings:

f. At the end of the process, you will be asked which settings you want to use: Last Used (L),

Recovery (R) or default (D). Type “L” and press RETURN.

This will use the previous settings loaded on the HygroGen2 irrespective of whether you performed

the clone process in step ii (b).

g. Acknowledge the warning. The unit will then shut down.

h. Remove the USB drive and turn the HygroGen2 back on. The unit will then boot into

CONFIGURATION MODE.

i. Check the Touch Screen Calibration and Date/Time and then press OK (or cancel to skip this)

to reboot the machine into OPERATIONAL MODE.

The instrument will now run using HygroGen2 Standard version 2.0 software.