Extremes of Temperature and Humidity

As with any other reference or instrument under test that is inserted through the door, stem conduction

should be considered at the extremes of the HygroGen2’s measurement range.

A recommendation for conditions that are higher than 50°C, 90%rh, or 60°C, 80%rh is that the entire

RP2 measuring head is mounted on the inside of the door. At these extremes, condensation will form at

cold spots that can be introduced if there is thermal conduction with the ambient conditions.

Figure 28: Probe mounted on inside door surface

An HG2-B8-9 bung which has a 9-17mm cable gland can be used to secure the head cable from the

outside of the door and provides further insulation.