4.2.3 Corrections

When getting your dew point hygrometer calibrated, please request a dew point calibration with

corrections from your calibration supplier and a table of corrections for applied set-points for your

external temperature probe. Enter the correction values from the instrument’s calibration certificate

via the two tables found via the “Corrections” option in the “External” drop-down menu and press the

toggle button so that it displays “Applied”.

Figure 26: External Reference Corrections

4.2.4 Calibration Info

Information about the calibration of your external reference can be entered via the “Calibration Info”

option in the “External” drop-down menu.

Figure 27: Reference Calibration Information