D1: HygroGen2 Uncertainty Framework - Internal RH Probe

Input values:

23

°C

80

% RH

Temperature sensitivity

4.96 % RH/deg C

Reference Uncertainty source

Distribution

Value

Units

Sensitivity

Equiv RH Divisor One SD Squares

1

Uncertainty of reference probe

normal

0.900000

%rh

1.000000

0.900000

2.00

0.450

0.203

2

Repeatability of humidity

normal

0.100000

%rh

1.000000

0.100000

1.00

0.100

0.010

3

Reference drift

rectangular 1.000000

%rh

1.000000

1.000000

1.73

0.577

0.333

4

Reference resolution

rectangular 0.050000

%rh

1.000000

0.050000

1.73

0.029

0.001

5

Chamber temperature gradients

rectangular 0.020000

°C

4.960000

0.099200

1.73

0.057

0.003

6

Temperature stabilisation

rectangular 0.005000

°C

4.960000

0.024800

1.73

0.014

0.000

7

Repeatability of temperature

normal

0.050000

°C

4.960000

0.248000

1.00

0.248

0.062

8

Humidity stabilisation

rectangular 0.050000

%rh

1.000000

0.050000

1.73

0.029

0.001

9

Humidity fluctuations

rectangular 0.050000 % RH

1.000000

0.050000

1.73

0.029

0.001

10

Instrument Under Test (IUT) Resolution

rectangular 0.050000 % RH

1.000000

0.050000

1.00

0.050

0.003

11

Standard deviation of Ref readings

normal

0.050000

%rh

1.000000

0.050000

1.00

0.050

0.003

12

Standard deviation of IUT readings

normal

0.100000

%rh

1.000000

0.100000

1.00

0.100

0.010

Reference

Sum of squares

0.63

1

From calibration certificate - check uncertainty at input

values

Standard uncertainty (√

sum of

squares)

0.793

2

From Standard Deviation of 10 repeated calibration runs

Expanded uncertainty (95 %

confidence)

1.59

3

Drift between two successive calibrations of the reference

instrument

4

Resolution from manufacturer’s specification

5

Measured temperature gradients from the ROTRONIC

HygroGen2 Service

6

Measured stability from the ROTRONIC HygroGen2 Service

7

From Standard Deviation of 10 repeated calibration runs

8

Resolution of ROTRONIC HygroGen2 Control Clip

9

Resolution of ROTRONIC HygroGen2 Control Clip

10

Resolution from manufacturer’s specification

11

Standard Deviation of Ref readings during calibration

12

Standard Deviation of IUT readings during calibration