Step 4.
Press the SCAN/adj key to display “OFFST”.

Step 5.
To calibrate voltage in (transducer), press the SCAN/adj key to display
“CAL U”. (If calibrating for voltage input, proceed to Step 7, if not, proceed to
Step 6 to calibrate for current input.)

Step 6.
To calibrate current in (transmitter), press the SCAN/adj key again to
display “CAL_C”, then proceed to Step 7.

Step 7.
Setting the Decimal Point.

Note: The decimal point location set in this step will always be used for all
displayed pressure readings on this channel. The maximum read-out is
+31999 to -9999.

Press the HOLD/mode key to display “SETdP”.

Step 7a.
Press and hold the SCAN/adj key and the decimal point will move to the left,
release key at desired location. (While the key is pressed, the decimal will
continue moving, left to right.) Press and release the SCAN/adj key to move
the decimal point one place at a time.