9

SET “OFFSET” (please read before attempting procedure)

Note: The channel must be calibrated before the offset is adjusted.

The ”OFFSET” command allows you to adjust the reading relative to a known
value. To set the ”OFFSET” use the following procedure:

Step 1.
Press and hold the SCAN/adj and HOLD/mode keys for 8 seconds, then
release; the display will alternate between “cal” and “noYes” for ten seconds.

Step 2.
To specify “yes” and select “CH1”, press and hold the HOLD/mode key to
display “CH1” . (To select “CH2”: With “CH1” displayed, press and release
the SCAN/adj key to display “CH2”.)

Step 3.
Press the HOLD/mode key to display “ALAr”.

Step 4.
Press the SCAN/adj key to display “OFFST”.

SCAN

HOLD

adj

mode

noYES

CH1

CH1

SCAN

HOLD

adj

mode

ALAr

SCAN

HOLD

adj

mode

OFF57

ALAr

0.00

CAL