Figure A

TRANSDUCERS ( 4 WIRE OR 4 TERMINAL DEVICES)

Connect the + excitation to pin 5 (CH2: pin 11) to supply 24 volts to the
transducer. IF YOUR TRANSDUCER REQUIRES 12 VOLT EXCITATION,
CONNECT THE + EXCITATION TO PIN 10 (CH2: PIN 10).

Connect the - excitation to Ground, pin 9 (CH2: pin 15).

Connect the + output to +IN, pin 6* (CH2: pin 12).

Connect the - output to -IN, pin 7* (CH2: pin 13).

Connect the shield to Ground, pin 9 (CH2: pin 15)

*NOTE: THE OUTPUT OF THE SETRA TRANSDUCER INSTALLED IN
THE DATUM 2000™ MANOMETER IS AVAILABLE ON PIN 6 AND 7. DO
NOT APPLY AN ELECTRICAL SIGNAL TO CH1: PIN 6 or CH1: PIN 7. A
SECOND TRANSDUCER MAY BE CONNECTED ONLY TO CH2: PIN 12
or CH2: PIN 13.

Supplied with
connector to join
the two foot
transducer cable
and the ten foot
digital pressure
readout cable

Setra pressure/vacuum
Transmitter

Pressure Line
(installed by user)

238652 REV B

DANGER

REFERENCE

VOLTAGE

SEE CALIBRATION

INSTRUCTIONS BEFORE

USING REF. VOLTAGE

>
>
>

>
>
>

Channel 1

Inputs/Excitations

Channel 2

Inputs/Excitations

Alarm

Signals

AC Adapter