Runtime:switch:delay, Default, Syntax – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 122: Example, Runtime:switch:delay -12
Advertising

7-12
85330A Multiple Channel Controller
Remote Programming Command Reference
Standard Command Reference
RUNTime:SWITch:DELay
Used for the delay between asserting the switch address list entry of the
RUNT:SWIT:SCAN command and the issuance of a receiver trigger. The
input to the command is in microseconds.
Default
The default is 2 microseconds.
Syntax
RUNT:SWIT:DEL <microseconds>;
Example
RUNT:SWIT:DEL 50;
! Set the trigger or setting
! time to 50 micro-seconds.
NOTE
•
When using the multiple-parameter display mode of the 8530A, the
minimum switch settling time setting is 50
µ
s. Using faster settling times
can cause problems during the measurement.
•
For the Fast Data Acquisition modes, 2
µ
s settling time is recommended.