Using imm vs. ttl trigger for source 1 – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 92

6-18
85330A Multiple Channel Controller
Programming
To Use Run-Time Control Mode
Run-Time
Measurement
Sequence for Multiple
Controllers
The run-time measurement sequence for multiple controllers is similar to a
single controller shown earlier. See
“Starting Run-Time Mode” on page 6-14
for more information on this sequence. The sequence for multiple
controllers has additional steps to allow communications between the
controllers. Refer to
for the actual run-time measurement sequence.
Using IMM vs. TTL
Trigger for Source 1
Refer to “Source 1 IMM, TTL or REM1 trigger diamond in
When using multi frequencies:
•
The RF Source controlled by the master 85330A should use the TTL
trigger.
•
The other RF Source (not controlled by the master 85330A) should use
the IMM trigger.