System:error, Default, Syntax – Agilent Technologies Multiple Channel Controller 85330A User Manual
Page 127: Example, System:error? -17, In case of difficulty
Advertising

85330A Multiple Channel Controller
7-17
Remote Programming Command Reference
Standard Command Reference
SYSTem:ERRor?
Allows checking the error status of the 85330A.
Default
The default should be: +0, “No error”.
Syntax
SYST:ERR?
Example
10 DIM ERR_STR$[128]
20 OUTPUT 70918;”SYST:ERR?;”
30 ENTER 70918;ERR_STR$
NOTE
The errors are put into a stack; therefore, reading the error status once may
not clear all error conditions. It should be read until the error +0,
“No error”
is
read. The
*CLS
command will clear the error status.