Pulse dose mode and autosat features, Warning – AirSep 4111-7 User Manual
Page 20

20
Personal Ambulatory Oxygen System
Users Manual
PULSE DOSE MODE AND autoSAT FEATURES
Your Eclipse has a feature, called autoSAT
®
that delivers a consistent-sized bolus of oxygen up
to 40 breaths per minute (see max breath rate chart below). As your breath rate increases, ie.
during ambulation, the autoSAT feature servo-controls the device delivering the set bolus size.
autoSAT Technology is proprietary to SeQual and is utilized during Pulse Mode only.
NOTE: Bolus volume decreases as breath rate exceeds published range.
BOLUS VOLUME (+/- 15%)
Pulse
Dose
Setting
Bolus Size
(± 15%) mL
AC Power Supply
and Power
Cartridge (Battery)
DC Power
Supply
Max Breath
Rate
Max Breath
Rate
1.0
16
40
40
2.0
32
40
40
3.0
48
40
40
4.0
64
40
31
5.0
80
37
25
6.0
96
31
20
-
128
23
15
-
160
18
12
-
192
15
10
Using the
Flow Setting Buttons, adjust the flow setting to the prescribed setting supplied by
your physician.
Step 5: ADJUST THE FLOW SETTING TO THE PRESCRIBED LEVEL
DO NOT USE SUPPLY TUBING OR CANNULA EXTENSIONS THAT ARE MORE THAN 50’ (15.2M) IN LENGTH
IN THE CONTINUOUS FLOW MODE.
PULSE DOSE MODE SETTINGS MUST BE DETERMINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR
NEEDS AT REST, DURING EXERCISE AND WHEN TRAVELING.
DO NOT EXCEED 7 FEET (2.1M) IN LENGTH OF SUPPLY TUBING OR CANNULA TUBING WHEN OPERATING
IN PULSE DOSE MODE IN ORDER TO ADEQUATELY DETECT ADEQUATE INSPIRATORY EFFORT.
AS WITH ALL OXYGEN-CONSERVING TYPE DEVICES, THE ECLIPSE MAY NOT BE ABLE TO DETECT ALL
INSPIRATORY EFFORTS IN PULSE MODE. IF THE ECLIPSE DOES NOT RESPOND TO YOUR INSPIRATION
EFFORTS, ASK YOUR HOME CARE PROVIDER TO CHECK THE SENSITIVITY SETTING.
DO NOT CHANGE FROM PRESCRIBED FLOW SETTING WITHOUT CONSULTING WITH A QUALIFIED
CLINICIAN OR PHYSICIAN.
WARNING: