Personal Ambulatory Oxygen System

Users Manual

PULSE DOSE MODE AND autoSAT FEATURES

Your Eclipse has a feature, called autoSAT

®

that delivers a consistent-sized bolus of oxygen up

to 40 breaths per minute (see max breath rate chart below). As your breath rate increases, ie.
during ambulation, the autoSAT feature servo-controls the device delivering the set bolus size.
autoSAT Technology is proprietary to SeQual and is utilized during Pulse Mode only.

NOTE: Bolus volume decreases as breath rate exceeds published range.

BOLUS VOLUME (+/- 15%)

Pulse

Dose

Setting

Bolus Size

(± 15%) mL

AC Power Supply

and Power

Cartridge (Battery)

DC Power

Supply

Max Breath

Rate

Max Breath

Rate

1.0

16

40

40

2.0

32

40

40

3.0

48

40

40

4.0

64

40

31

5.0

80

37

25

6.0

96

31

20

-

128

23

15

-

160

18

12

-

192

15

10

Using the

Flow Setting Buttons, adjust the flow setting to the prescribed setting supplied by

your physician.

Step 5: ADJUST THE FLOW SETTING TO THE PRESCRIBED LEVEL

DO NOT USE SUPPLY TUBING OR CANNULA EXTENSIONS THAT ARE MORE THAN 50’ (15.2M) IN LENGTH
IN THE CONTINUOUS FLOW MODE.

PULSE DOSE MODE SETTINGS MUST BE DETERMINED FOR EACH PATIENT INDIVIDUALLY FOR THEIR
NEEDS AT REST, DURING EXERCISE AND WHEN TRAVELING.

DO NOT EXCEED 7 FEET (2.1M) IN LENGTH OF SUPPLY TUBING OR CANNULA TUBING WHEN OPERATING
IN PULSE DOSE MODE IN ORDER TO ADEQUATELY DETECT ADEQUATE INSPIRATORY EFFORT.

AS WITH ALL OXYGEN-CONSERVING TYPE DEVICES, THE ECLIPSE MAY NOT BE ABLE TO DETECT ALL
INSPIRATORY EFFORTS IN PULSE MODE. IF THE ECLIPSE DOES NOT RESPOND TO YOUR INSPIRATION
EFFORTS, ASK YOUR HOME CARE PROVIDER TO CHECK THE SENSITIVITY SETTING.

DO NOT CHANGE FROM PRESCRIBED FLOW SETTING WITHOUT CONSULTING WITH A QUALIFIED
CLINICIAN OR PHYSICIAN.