Troubleshooting table, Warning – AirSep 4111-7 User Manual

Page 22

Advertising
background image

22

Personal Ambulatory Oxygen System

Users Manual

No Battery installed

Install

Battery

or plug into external power

Battery is discharged

or no external power is present

Plug into external power

AC or DC Power Supply

is not plugged in correctly

Check plug at outlet, Power Supply box and

at the Eclipse to ensure properly plugged in.

Insufficient power from the outlet

Ensure there is adequate power from the outlet.

(Outlet may be controlled by a wall switch

or the circuit breaker may have engaged.)

Other

Contact your dealer/provider

Filter Blocked

Clean Air Inlet Filter

Humidifier

Check humidifier attachment

and tubing

Eclipse not ON

Power Eclipse ON

Tubing or cannula is not properly

connected or kinked

Check tubing, cannula

and connections

Other

Contact your dealer/provider

Restriction in tubing or humidifier

Repair or replace tubing or humidifier.

Place your Eclipse so there is adequate air flow.

Filter restricted

Clean or replace air inlet filter. Place your Eclipse

so there is adequate air flow.

Exceeding maximum breath rate in

Pulse Mode. *See Pulse Dose Mode

Breath Rate Chart (p. 18)

Switch to Continuous Flow Mode until respiratory

rate resumes to an acceptable rate.

Other

Contact your dealer/provider

Restriction in humidifier or tubing

Repair or replace humidifier or tubing

Filter Blocked

Clean or replace air inlet filter

Other

Contact your dealer/provider

Tubing/cannula longer

than 7 feet (2.1m)

Attach 7 foot (2.1m) tubing/cannula

Humidifier attached

Remove humidifier

No inspiration detected

Contact your dealer/provider

Unit in Continuous Mode

Switch to Pulse Mode

Nasal cannula not in the

nose or mouth breathing

Place nasal cannula in nostrils

and breath through your nose

Power Cartridge (Battery) is aging

Contact your dealer/provider

to replace Power Cartridge (Battery)

SYMPTOM

PROBABLE CAUSE

REMEDY

Eclipse does not

power on when

ON/OFF button

is pressed

No Oxygen

Delivered in

Pulse Flow Mode

Power Cartridge Status

Gauge never indicates

fully charged

Low Oxygen

Concentration

Low Oxygen Flow

No Oxygen

TROUBLESHOOTING TABLE

DO NOT IGNORE ALARMS.

A BACK-UP SOURCE OF SUPPLEMENTAL OXYGEN IS RECOMMENDED IN CASE OF A POWER OUTAGE
OR A MECHANICAL FAILURE OF THE DEVICE. CONSULT YOUR PROVIDER FOR A BACK-UP SOURCE
OF OXYGEN.

WARNING: