Boonton 9240 Series RF Voltmeter

CALCulate:MATH:ARGA

Description:

Set or return the first argument to be used for channel math operations.


Syntax:

CALCulate:MATH:ARGA <character data>


Argument:

<character data> = { CH1, CH2 }



CALCulate:MATH:ARGB

Description:

Set or return the second argument to be used for channel math operations.


Syntax:

CALCulate:MATH:ARGB <character data>


Argument:

<character data> = { CH1, CH2 }



CALCulate:MATH:DATA?

Description:

Returns the channel math result.


Syntax:

CALCulate:MATH:DATA?



CALCulate:MATH:OPERator

Description:

Set or return the channel math operator.


Syntax:

CALCulate:MATH:OPERator <character data>


Argument:

<character data> = { CH_SUM, CH_DIFF, CH_RAT }




CALCulate:MODe

Description:

Set or return the system remote measurement mode. These measurement modes are only available
under bus control. NORMal is the default mode of operation at approximately 20 readings per
second. FAST returns data at over 200 readings per second. FILTered will only return data when
the filter is full. FILTered works best when INITiate:CONTinuous is OFF and a trigger is issued.
In this case the measurement will not be returned until the filter is full. When the instrument is
returned to local the MODE defaults to NORMal.


Syntax:

CALCulate:MODe <character data>


Argument:

<character data> = { NORMal, FAST, FILTered }

Remote Operation

5-12