12 status commands, Status:device:condition, Status:device:enable – Boonton 9240 RF Voltmeter User Manual

Page 74: 12 status commands -24, Status:device:condition? -24, Status:device:enable -24, Boonton 9240 series rf voltmeter, Remote operation 5-24

Advertising
background image

Boonton 9240 Series RF Voltmeter

5.5.12 STATus Commands

The STATus command subsystem enables you to control the SCPI defined status reporting structures. The user may
examine the status or control status reporting of the voltmeter by accessing the Device, Operation and Questionable status
groups

.

STATus:DEVice:CONDition?

Description:

Return the current value of the Device Condition register. The following table shows the bit
assignments in the register. These bits are updated by the instrument in real time, and can change
in response to changes in the instrument’s operating condition.

Bit

Value

Definition

0

1

Not used

always returns 0.

1

2

Channel 1 Connected

1 = A sensor or probe is connected to channel 1.

2

4

Channel 2 Connected

1 = A sensor or probe is connected to channel 2.

3

8

Channel 1 Error

1 = Channel 1 is reporting an error.

4

16

Channel 2 Error

1 = Channel 2 is reporting an error.

5

32

Shape Cal 1

1 = Channel 1 is using a CW shape cal table.

6

64

Shape Cal 2

1 = Channel 2 is using a CW shape cal table.

7

128

Not used

always returns 0.

8

256

Not used

always returns 0.

9

512

Not used

always returns 0.

10

1024

Not used

always returns 0.

11

2048

Not used

always returns 0.

12

4096

Not used

always returns 0.

13

8192

Key Press

1 = A key has been pressed.

14

16384

Not used

always returns 0.

15

32768

Not used

always returns 0.


Syntax:

STATus:DEVice:CONDition?

Returns:

16-bit register value (0 to 65535)

STATus:DEVice:ENABle

Description:

Sets or returns the Device Enable register, which contains the bit mask that defines which true
conditions in the Device Status Event register will be reported in the Device Summary bit of the
instrument Status Byte. If any bit is 1 in the Device Enable register and its corresponding Device
Event bit is true, the Device Status summary bit will be set.

Syntax:

STATus:DEVice:ENABle <numeric value>

Argument:

<numeric value> = 0 to 65535

Returns:

16-bit register value (0 to 65535)

Remote Operation

5-24