Probe replacement – Sper Scientific 840034 Psychrometer - Anemometer User Manual

Page 20

Advertising
background image

- 20 -

c. If the S value is 0.2396838, you should enter

00 23968.

For Intercept (value Z)

1. After pressing

Enter to save the S setting, the

meter will advance to the intercept (Z) input
setting automatically. The last digit will flash to
indicate the Z value is now ready for editing.

2. If the last saved Z value is different from the slip of

paper, press

Hold to move to the value you want

to edit and then use

Mx/Mn to change the value.

3. Press

Enter to save the new Z value. The meter

will return to the normal RH value mode.

Note: While setting the Z value, only input three dig-
its after the decimal point. You will need to round off
the fourth digit.

Examples:

a. If the Z value is -928.8683, you should enter

09 28 868.

b. If the Z value is -1928.868, you should enter

19 28 868.

c. If the Z value is -89.68482, you should enter

00 89 685.

PROBE REPLACEMENT