Probe replacement – Sper Scientific 840034 Psychrometer - Anemometer User Manual
Page 20
- 20 -
c. If the S value is 0.2396838, you should enter
00 23968.
For Intercept (value Z)
1. After pressing
Enter to save the S setting, the
meter will advance to the intercept (Z) input
setting automatically. The last digit will flash to
indicate the Z value is now ready for editing.
2. If the last saved Z value is different from the slip of
paper, press
Hold to move to the value you want
to edit and then use
Mx/Mn to change the value.
3. Press
Enter to save the new Z value. The meter
will return to the normal RH value mode.
Note: While setting the Z value, only input three dig-
its after the decimal point. You will need to round off
the fourth digit.
Examples:
a. If the Z value is -928.8683, you should enter
09 28 868.
b. If the Z value is -1928.868, you should enter
19 28 868.
c. If the Z value is -89.68482, you should enter
00 89 685.
PROBE REPLACEMENT