MODE AND UNIT SELECTION

Zero Adjustment

1. Press MODE continuously to cycle through

modes until you reach Zero Mode.

2. Cover the sensor. If the main display has a

reading other than zero, perform the zero
adjustment.

3. Cover the sensor and press S to zero adjust.

The secondary display indicates the
adjustment value for your reference.

4. Press and hold S for 1 second to clear the

adjustment value.

Note…
When the adjustment value is over 10

W, the

measurement value cannot be offset. The main
display will show the above range reading, and
the secondary display will show “-Err.” Repeat
the zero adjustment procedure or contact Sper
Scientific.

MODE AND UNIT SELECTION

5. In Zero Mode, press M to clear data points 0

to 20.

Unit Selection

1. Press MODE continuously to cycle through

modes until you reach Unit Mode.

2. Press S to choose the unit of measure:

W/

cm

2

or

mW/cm

2

.

Type Selection

This function is reserved for use in future
models.