Agilent Technologies N8201A User Manual

Page 50

Advertising
background image

46

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

2

Hardware Setup and Configuration

Measurement Procedure

5

Tune the Spectrum Analyzer to the following:

•

Frequency: 7.5 MHz

•

Amplitude: 10 dBm

•

Span: 100 kHz

6

Tune the downconverter to the following:

•

Level: Attenuator: 20 dBm

•

RF Enabled: Enabled

•

Input RF: Signal Source: 50 MHz Cal

•

Frequency: 50 MHz

7

Verify that a 7.5 MHz signal is present on the spectrum analyzer.