Vaisala Calibration of Digital Transmitters with HMI41 User Manual

Page 11

Advertising
background image

Chapter 2 __________________________________________________________ Product Overview

VAISALA ________________________________________________________________________ 9

set

The number on the first line of the display corresponds to the
following HMI41 functions:

1

=

HMI41 as a humidity and temperature indicator

2

=

HMI41 as a calibrator for HMD/W60/70 and HMP140 series
analogue transmitters

3

=

HMI41 as a calibrator for HMP230 and HMT100/HMT330
series digital transmitters

4

=

HMI41 as a calibrator for the HMD/W20/30 and HMP130
series analogue transmitters

To calibrate HMP230, HMT100 or HMT330 series digital
transmitters, select number 3 with buttons ▲ (number up) and ▼
(number down) and then press ENTER. The following text appears:

set

If the baud rate on the HMI41 display is not the same as that of the
transmitter, change it with buttons ▲ and ▼. Use baud rate 19200 (=
19.2 on the HMI41 display) with the HMT100 and HMT330 series.
When the baud rate is the same, press ENTER and the following
appears:

set

Letters and numbers correspond to the following:

E = even

parity

7 = data

bits

1 = stop

bits

Advertising