Two-point calibration – Vaisala Calibration of Digital Transmitters with HMI41 User Manual

Page 20

Advertising
background image

User's Guide ______________________________________________________________________

18 __________________________________________________________________ M210777EN-B

In this case, perform the calibration again. Whether the calibration
was successful or not, HMI41 always returns to show the selected
calibration mode:

RH

or

RH

After having successfully completed the calibration, turn HMI41 off
and disconnect the cable.

Two-Point Calibration

In two-point calibration mode, both offset and gain corrections are
made. This calibration mode is more accurate than offset or gain
calibrations. For performing a two-point calibration you need two
reference points with a difference of at least 50 %RH between them.

NOTE

Two-point calibration is not supported for HMT100.


Select two-point calibration with buttons ▲ and ▼:

RH

Press ENTER and the following text appears on the display for about
1 ... 2 seconds:

RH