One-point offset and gain calibrations, Hmi41 as a reference meter – Vaisala Calibration of Digital Transmitters with HMI41 User Manual

Page 17

Advertising
background image

Chapter 3 _______________________________________________________________ Calibration

VAISALA _______________________________________________________________________ 15

One-Point Offset and Gain
Calibrations

Select the desired calibration with buttons ▲ and ▼. Select offset
calibration if the humidity is < 65%RH and gain calibration if the
humidity is > 65%RH:

RH

or

RH

Press ENTER to acknowledge the selection. A text similar to the
following appears on the display:

RH

%

In following chapters, you will find instructions for performing offset
and gain calibrations in two different ways.

HMI41 as a Reference Meter

The HMI41 reference probe is used to show the correct humidity
value. Connect the cable and wait for the transmitter and reference
probe readings to stabilize; when both readings are stable, press
ENTER to conclude the calibration. When ENTER is pressed, the
transmitter reading is corrected to the HMI41 reference probe reading.

RH

%

Numbers on the first line indicate the RH measured by the transmitter,
and numbers on the second line indicate the RH measured by the
reference probe. Let the readings stabilize.