Vaisala HMP140 User Manual

Page 17

Advertising
background image

HMP140A SERIES

U175en-1.5

Operating Manual

1999-12-08

13

6.3

General

Power supply
current outputs:

20...35 VDC

17...24 VAC

min.

12 +

R

L

(ohm)

50

0...1 V

9...35 VDC

9...24 VAC

0...5 V

15...35 VDC

12...24 VAC

0...10 V

20...35 VDC

16...24 VAC

Current consumption

6...10 mA + output currents

Electrical connections

screw terminals for 0.5 - 1.5 mm

2

wires (AWG

15...20), stranded wires recommended

Alternative analogue
output signals (scalable)

0...20 mA (4...20 mA)
0...1 V, 0...5 V, 0...10 V

NOTE! Output scales correspond to 0…100 %RH and -40…+60°C.

External loads for
current outputs

< 500

0...1 V

> 1 k

(to ground)

0...5 V

> 5 k

(to ground)

0...10 V

>10 k

(to ground)

Operating temperature range
for electronics

-20...+60

°

C

with display option

0...+50

°

C

Storage temperature range -40...+80

°

C

Operating humidity range

0...100 %RH

Bushing

for

7...10 mm cables

Sensor protection
standard

membrane filter, part no. 17039HM

option

plastic grid, part no. 17038

Cable length of HMP143A
standard

2.5 m

maximum

100 m (user extendable, see Appendix 3)

Housing material

ABS plastic

Housing classification

IP65 (NEMA 4)

Advertising