Filt set result filtering, Reset reset transmitter, Errs display error messages – Vaisala HMT310 User Manual
Page 50

USER'S GUIDE____________________________________________________________________
50 __________________________________________________________________ M210619EN-D
FILT Set Result Filtering
Syntax: FILT xx<cr>
Enable or disable the filtering or select the extended filter to reduce noise
of the measurement.
Where
x = ON, OFF or EXT
ON = Short filter of about 15 s (results the average value of the
last 15 s measurement data)
OFF = No filtering (default)
EXT = Extended filter of about 1 min (results the average value
of the last 1 min measurement data)
RESET Reset Transmitter
Syntax: RESET<cr>
ERRS Display Error Messages
Syntax: ERRS<cr>
Display transmitter error messages. If there are no errors present,
a PASS will be returned.
Examples:
>errs<cr>
PASS
>
>errs<cr>
FAIL
Error: Temperature measurement malfunction
Error: Humidity sensor open circuit
>