FCRH Relative Humidity Calibration after
Sensor Change

Syntax: FCRH<cr>

The transmitter asks and measures relative humidity and calculates the
calibration coefficients. This two-point adjustment needs to be performed
after a sensor change. Follow the more detailed calibration instructions
on page 69, but instead of the CRH command, use the FRCH command.

Example:

>frch<cr>
RH :

1.82

1.

ref

?

0<cr>

Press any key when ready<cr>
RH

:

74.22

2. ref

? 75<cr>

OK
>

The OK indicates that the calibration has succeeded.

CTEXT Set Calibration Information Text

Syntax: CTEXT<cr>

1.

Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustm
ent button once (see Figure 17 on page
70).

2.

Issue command LI and enter value 0 for an offset value and value 1
for a gain value.

3.

Issue command CTEXT and enter calibration information text.

Example:

>ctext<cr>
Cal. info : Vaisala/HEL ? HMK15<cr>
>

CDATE Set Calibration Date

Syntax: CDATE yyyy mm dd<cr>

1.

Remove the transmitter unit from the mounting plate (see
Mounting the Transmitter/Removing the Transmitter Unit on page
21) and press the adjustm
ent button once (see Figure 17 on page
70).