Chapter 7 ___________________________________________________Calibration and Adjustment

VAISALA _______________________________________________________________________ 69

2.

Issue command LI and enter value 0 for an offset value and value 1
for a gain value.

3.

Issue command CDATE and set the calibration date.

Example:

>cdate 2001 12 11<cr>
Calibration : 2001-12-11
>

ACAL Analog Output Calibration

Syntax: ACAL<cr>

Connect the HMT310 to a multimeter in order to measure either current
or voltage depending on the selected output type. Issue the ACAL
command. Then type the multimeter reading and press Enter.

Example (current outputs):

>acal<cr>
Ch1

I1

(mA) ?

2.046 <cr>

Ch1 I2

(mA)

? 18.087

<cr>

Ch2

I1

(mA) ?

2.036 <cr>

Ch2 I2

(mA)

? 18.071

<cr>

>

Relative Humidity Calibration and Adjustment
(in Two Points)

Use two reference humidities over the measurement range. The
references need to have a difference of at least 50 %RH.

Before calibration HMT310 needs to be set to adjustment mode by
pressing the adjustment button once, see Figure 17 on page 70. (To return
to normal operation mode, use command RESET, see RESET Reset
Transmitter on page 50).