Problem

Possible Cause

Solution

The parameter I want
is not selectable.

Connected instrument
does not provide the
desired parameter.

Check the instrument's
documentation to verify
the parameters that are
available.

Parameter on screen
shows stars "***"
instead of
measurement reading.

Measurement problem in
the connected instrument:
error or out of
measurement range.

Check and correct.

Measurement unit is
not the one I want.

Position of DIP switch 8 on
RDP100.

Set DIP 8 to metric or
non-metric as applicable.

RDP100 Maintenance

Calibration

RDP100 does not need calibration. If the measurements on screen are not
correct, check and calibrate the measurement instrument that is connected to
the RDP100.

Cleaning

Wipe the front cover with a soft, lint-free cloth moistened with an isopropyl
alcohol (IPA) solution or a mild detergent.

If there is dust on the component board and the display element, open the front
cover and clean the parts by gently blowing with instrument air. Avoid touching
the display element.

Spare Parts

Replacement front cover is available as spare part (Vaisala order code
ASM211018SP).