Chapter 8 _____________________________________________________________ Technical Data

VAISALA ________________________________________________________________________ 63

CHAPTER 8

TECHNICAL DATA

This chapter provides the technical data of Dewpoint Transmitter
DMT152.

Measured Variables

Dewpoint temperature

Measurement range

-80 … -10 °C (-112 … +14 °F) T

d

Accuracy

-80 … -30 °C (-112 … -22 °F)

± 2 °C (3.6 °F) T

d

-30 … -10 °C (-22 … +14 °F)

± 3 °C (5.4 °F) T

d

Non-calibrated range

-100 … -80 °C, -10 … +20 °C

(-148 … -112 °F

,

+14 … +68 °F T

d

)

Response time 63% [90%] at +20 °C (+68 °F)
gas temperature and pressure of 1 bar

-10 ... -80 °C (+14 … -112 °F) T

d

0.5 min [7.5 min]

-80 ... -10 °C (-112 … +14 °F) T

d

2 s [5 s]

Analog output scalings

option 1

-80 ... +20 °C (-112 ... +68 °F) Td

option 2

-100 ... 0 °C (-148 ... +32 °F) Td

option 3

free scaling

ppm volume concentration

Measurement range

0 … 500 ppm

Accuracy at +20 °C (+68 °F),
1013 mbar

± (0.2 ppm + 20 % of reading)

0904-007-B

Figure 23

Accuracy Over Temperature Range

When dewpoint is below 0 °C, the transmitter outputs frostpoint for Td.