List of Serial Commands

All commands can be issued either in uppercase or lowercase. In the

command examples, the keyboard input by the user is in bold type.

The notation <cr> refers to pressing the carriage return (Enter) key on

your computer keyboard. Enter a <cr> to clear the command buffer

before starting to enter commands.

Table 6

List of Serial Commands

Command

Description

?

Show device information

??

Show device information (overrides poll mode)

ADDR [0 … 255]

Show or set transmitter address

AERR

Set analog output error level

AMODE

Set analog output mode

AOVER [ON/OFF]

Allow analog outputs to exceed their range 10%

ASEL

Set analog output parameters and scaling

ATEST

Test analog outputs

CLOSE

Close the temporary connection and return the

device to POLL mode

ERRS

Display the currently active errors

FORM [modifier string] Set output format

FRESTORE

Restore factory settings

HELP

Show list of currently available serial commands

INTV [0 ... 255 S/MIN/H] Set the continuous output interval (for RUN mode

and R command)

OPEN [0 … 255]

Open a line to transmitter at defined address

R

Start the continuous outputting

RESET

Reset the transmitter

S

Stop the continuous outputting

SDELAY [0 … 255]

Set response sending delay

SEND [0 … 255]

Output readings once (specify address for

transmitters in POLL mode)

SERI [baud p d s]

Show or set the serial interface settings

SMODE [mode]

Show or set startup serial mode:

RUN, STOP, POLL, or MODBUS

SNUM

Display transmitter serial number

TIME

Show transmitter uptime

UNIT [m/n]

Select metric or non-metric units

VERS

Show transmitter firmware version