Regulatory compliances, Table 3, Environmental tests – Vaisala WMT700 User Manual

Page 17

Advertising
background image

Chapter 1 _________________________________________________________ General Information

Regulatory Compliances

Vaisala WINDCAP

®

Ultrasonic Wind Sensor WMT701,

WMT702, and WMT703 comply with the performance and
environmental test standards listed in Table 3 below.

Wind tunnel tests have been performed according to Sonic
anemometers/thermometers - Acceptance test methods for mean
wind measurements ISO 16622:2002 and Measnet Anemometer
Calibration Procedure Version 2, October 2009.

Table 3

Environmental Tests

Test

Setup According to

Wind driven rain

MIL-STD 810G Method 506.5 and
Telcordia GR-487-Core

Salt fog

VDA 621 - 415 / IEC 60068-2-52

Leak test (Ingression Protection)

IEC 60529 class IP67

Vibration

IEC 60068-2-6 / IEC 60945 /
Lloyd's test

Shock

MIL-STD-202G, Method 213B, cond. J

Dry heat

IEC 60068-2-2 / IEC 60068-2-48

Damp heat cyclic

IEC 60068-2-30, Test Db

Damp heat

IEC 60068-2-78

Low temperature

IEC 60068-2-1 Test Ab/Ad

Free fall (rough handling)

IEC 60068-2-31

Change of temperature

IEC 60068-2-14

EMC tests are based on a European product family standard:
EN 61326-1:2006 (Electrical equipment for measurement, control
and laboratory use - EMC requirements for use in industrial
locations) and EN 60945:2002 (Maritime Navigation and
Radiocommunication Equipment and Systems - General
Requirements - Methods of Testing and Required Test Results).

VAISALA ________________________________________________________________________ 15