Message

Typical

Min/
Max

Fault

Description

Action

P15
Positive voltages
of the DC / DC
converter for the
transmitter/receiver
.

+15.0 V

+14 V/
+16 V

<+14.0 V
>+16.0 V

DC/DC converter is
overloaded or working
wrong.

1. Check flat cable
between
transmitter/
receiver.
2. Change the
transmitter.
3. Change the
receiver.

M15
Negative voltages
of the DC/DC
converter for the
transmitter/receiv
er

-15.0 V

-16 V/
-14 V

>-14.0 V
<-16.0 V

DC/DC converter is
overloaded or working
wrong.

1. Check flat cable
between
transmitter/
receiver.
2. Change the
transmitter.
3. Change the
receiver.

BGDN
Ground potential
for the transducer,
i.e.,

voltage loss in

transducer cable.

~0 V

-0.5 V/
+1 V

<-0.5 V
>+1 V

Check transducer
cable.

Getting Help

For technical support or comments on the manual, contact the Vaisala
technical support:

E-mail

[email protected]

Telephone

+358 9 8949 2789

Fax

+358 9 8949 2790

Return Instructions

If the product needs repair, please follow the instructions below to
speed up the process and avoid extra costs.

1.

Read the warranty information.

2.

Write a Problem Report with the name and contact information
of a technically competent person who can provide further
information on the problem.