Performance characteristics, Continued) – Agilent Technologies InfiniiVision 6000 Series User Manual
Page 19

19
Performance characteristics
(continued)
Trigger system (continued)
Selections
Edge, pulse width, pattern, TV, duration, sequence, CAN, LIN, USB, I
2
C, SPI, Nth edge burst
Edge
Trigger on a rising, falling, alternating or either edge of any source
Pattern
Trigger at the beginning of a pattern of high, low, and don’t care levels and/or a rising or
falling edge established across any of the analog and digital channels, but only after a
pattern has stabilized for a minimum of 2 nsec.
The scope channel’s high or low level is defined by that channel’s trigger level. The logic
channel’s trigger level is defined by the threshold for the pod, 0 - 7 or 8 - 15.
Pulse width
Trigger when a positive- or negative-going pulse is less than, greater than, or within a
specified range on any of the source channels.
Minimum pulse width setting: 5 ns (MSO/DSO601xA/603xA scope channels)
2 ns (MSO/DSO605xA/610xA scope channels)
2 ns (logic channels on MSO6000A or
MSO-upgraded DSO6000A)
Maximum pulse width setting: 10 s
TV
Trigger using any scope channel on most analog progressive and interlaced video standards
including HDTV/EDTV, NTSC, PAL, PAL-M or SECAM broadcast standards. Select either
positive or negative sync pulse polarity. Modes supported include Field 1, Field 2, all fields,
all lines, or any line within a field. TV trigger sensitivity: 0.5 division of sync signal. Trigger
holdoff time can be adjusted in half field increments.
Sequence
Arm on event A, trigger on event B, with option to reset on event C or time delay.
CAN
Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. Trigger on the start
of frame (SOF) bit (standard). N5424A option supports triggering on remote frame ID (RTR),
data frame ID (~RTR), remote or data frame ID, data frame ID and data, error frame, all errors,
acknowledge error and overload frame.
LIN
Trigger on LIN (Local Interconnect Network) sync break at beginning of message frame
(standard). N5424A option supports triggering on frame ID.
USB
Trigger on USB (Universal Serial Bus) start of packet, end of packet, reset complete, enter
suspend, or exit suspend on the differential USB data lines. USB low speed and full speed
are supported.
I
2
C
Trigger on I
2
C (Inter-IC bus) serial protocol at a start/stop condition or user defined frame
with address and/or data values. Also trigger on missing acknowledge, address with no acq,
restart, EEPROM read, and 10-bit write.
SPI
Trigger on SPI (Serial Protocol Interface) data pattern during a specific framing period.
Supports positive and negative Chip Select framing as well as clock Idle framing and
user-specified number of bits per frame.
Duration
Trigger on a multi-channel pattern whose time duration is less than a value, greater than a
value, greater than a time value with a timeout, or inside or outside of a set of time values.
Minimum duration setting: 2 ns
Maximum duration setting: 10 s
Nth edge burst
Trigger on the Nth edge of a burst that occurs after an idle time that you specify. Max edge
count: 65,536.
Autoscale
Finds and displays all active scope and logic (for MSO6000A series MSO) channels, sets edge
trigger mode on highest-numbered channel, sets vertical sensitivity on scope channels and
thresholds on logic channels, time base to display ~1.8 periods. Requires minimum voltage
>10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.