Scope channel triggering
Range (internal)
±6 div from center screen
Sensitivity*
<10 mV/div: greater of 1 div or 5 mV;
≥10 mV/div: 0.6 div
Coupling
AC (~3.5 Hz on MSO/DSO601xA, ~10 Hz on MSO/DSO603xA/605xA/610xA),
DC, noise reject, HF reject and LF reject (~50 kHz)
Logic (D15 - D0) channel triggering (MSO6000A or MSO-upgraded DSO6000A only)
Threshold range (user defined)
±8.0 V in 10 mV increments
Threshold accuracy*
±(100 mV + 3% of threshold setting)
Predefined thresholds
TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
External (EXT) triggering
MSO/DSO6xx2A (2-/2+16-ch models)
MSO/DSO6xx4A (4-/4+16-ch models)
Input impedance
MSO/DSO6012A: 1 M
Ω ± 3% || 11 pF or 50 Ω
MSO/DSO6014A: 1.015 k
Ω ±5%
MSO/DSO6032A/6052A/6102A: MSO/DSO6034A/6054A/6104A:
1 M
Ω ± 3% || 14 pF or 50 Ω
2.14 k
Ω ±5%
Maximum input
CAT I 300 Vrms, 400 Vpk, CAT II 100 Vrms, 400 Vpk
±15 V
With 10073C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
5 Vrms with 50-ohm input
Range
DC coupling: trigger level ±1 V and ±8 V
±5 V
Sensitivity
For ±1 V range setting: DC to 100 MHz, 100 mV;
MSO/DSO6014A:
MSO/DSO6032A/6052A/6102A:
DC to 100 MHz: 500 mV
>100 MHz to bandwidth of oscilloscope: 200 mV
MSO/DSO6034A/6054A/6104A:
For ±8 V range setting: DC to 100 MHz, 250 mV;
DC to 500 MHz: 500 mV
MSO/DSO6032A/6052A/6102A:
>100 MHz to bandwidth of oscilloscope: 500 mV
Coupling
AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~50 kHz)
Probe ID
MSO/DSO601xA: Auto probe sense
MSO/DSO603xA/605xA/610xA: Auto probe sense and AutoProbe interface
Agilent- and Tektronix-compatible passive probe sense
Display system
Display
6.3-inch (161 mm) diagonal color TFT LCD
Throughput of scope channels
Up to 100,000 waveforms/sec in real-time mode
Resolution
XGA – 768 vertical by 1024 horizontal points (screen area);
640 vertical by 1000 horizontal points (waveform area)
256 levels of intensity scale
Controls
Waveform intensity on front panel. Vectors on/off; infinite persistence on/off,
8 x 10 grid with intensity control
Built-in help system
Key-specific help (in English) displayed by pressing and holding key or softkey of interest
Real-time clock
Time and date (user adjustable)
*
Denotes warranted specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.