Journal interval – Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 62

Advertising
background image

60

Set up Journal Mode

Journal mode is similar to Automatic Datalogging mode where

the operator can set a timed interval between data points being

communicated to the PC.
In Journal Mode the HHC will output the data string at a time

interval set by the user. No command from the PC is needed to

start the Journal output. The Journal mode output is started

and stopped via the user from the keypad of the HHC
Press MENU key

Use DOWN or UP ARROW to highlight “Modes…”

Press ENTER

Use DOWN or UP ARROW to highlight “Communications …”

Press ENTER

Use DOWN or UP ARROW to highlight “Journal/Inquiry…”

Press ENTER

Use DOWN or UP ARROW to highlight “Journal Interval…”

Press ENTER

Display will show “Journal interval”

Note: maximum journal interval is 86400 seconds (24 hours)
Press left and RIGHT ARROWs to select desired numeric

character.
Press UP and DOWN keys to increment values UP or DOWN.
When the desired interval is displayed Press ENTER to save.
*NOTE*

To begin or stop detecting journal data when in Journal/Inquiry

menu use DOWN or UP ARROW to hilight “Journal Start/Stop”.

Press ENTER. When programming, the display will show

“Desired” interval and “Actual” interval. Actual interval may not

match the desired value. The minimum sample rate changes are

based on the number of inputs installed in the base unit. The

sample rate increase 0.1 seconds per input (0.1, 0.2 or 0.3 sec-

onds). The actual interval must be an even multiple of this value.

For example if there are two pressure modules and a voltage

input active, the minimum sample rate is 0.3 seconds or even

multiples of 0.3 seconds. If the desired value entered is 1 second

the actual value will remain 1.2 seconds.

ISO1745

Refer to section 16c for protocol definition and commands.

HHC Base Unit Field Calibration

The HHC base unit can be recalibrated and certified in the field.

The procedure allows for the acquisition of “as found” and “as left”

data, and recalibration of the base unit.

Field calibration typically consists of three steps. Step one is to

document the “as found” or pre calibration condition of the HHC.

Step two is to perform the recalibration procedure to adjust coeffi-

cients used by the HHC. Step 3 is to repeat step one to document

the “as left” or post recalibration condition of the HHC.

The procedure details three processes, first obtaining “as found”

and second performing the actual calibration of the HHC, and third

obtaining “as left” data..

SECTION 15 m. v.

Journal Mode

SECTION 15 m. vi.

ISO 1745 Mode

SECTION 15 n.

CALIBRATE BASE UNIT

Journal interval

Desired: 00001 sec

Actual: 00001.2 sec

Enter to accept

Esc to cancel

Advertising