Ashcroft ATE-2 - Handheld Calibrator User Manual

Page 78

Advertising
background image

76

Danger, the PTE-2 may not display the full scale psi when

applied (A legacy PTE-1 module with invalid output at full

scale) Contact Customer Service

•

CH1 Low Span / CH2 Low Span

Danger, the PTE-2 may not display the full scale psi when

applied (A legacy PTE-1 module with invalid output at full

scale) Contact Customer Service

• Loop supply overloaded: see troubleshooting section
• Loop supply reset failed: see troubleshooting section
• FRAM write failed: Contact customer service
• FRAM read failed: Contact customer service
• Datalog corrupt: Contact customer service
•

Datalog error corrected: if warning reoccurs contact customer

service

•

Can’t read clock: There is an issue with the internal real time

clock,

• Current date bad: the real time used may not be accurate.
•

Failed to open file/ Failed to close file: SD card failure, card

may be bad or full

System Warnings

• SYS not-calibrated: The PTE-2 is not calibrated, contact cus-

tomer service

•

Out of internal memory: datalog memory is filled, transfer to

SD Card and clear internal datalog to make room.

•

Out of datalog: maximum number of datalogs reached, trans-

fer to SD Card and clear internal datalog to make room.

•

Out of records: present datalog can take no more records,

start a new datalog.

Operation Warnings

•

CH1 zero out of bounds/ CH2 zero out of bounds: input pres-

sure is too high to be zeroed

•

CH1 not a pressure mod / CH2 not a pressure mod: attempt-

ing to zero or tare a module which is not a pressure module

•

CH1 conversion invalid/ CH2 conversion invalid: module

range is not valid for Units selection

•

No SD card present: Attempting to perform an SD card oper-

ation with non-installed SD card

•

SD card locked: attempting to transfer to an SD card but the

write protect tab is engaged.

•

No module to damp: attempting to damp with no module

present

•

Can’t enter MENU (log): The MENU cannot be entered while

datalogging, press the Esc key to halt a datalog when ready.

System Warnings

• Low battery: main battery is low, replace with four fresh AA

alkaline batteries

•

RTC battery low: Real time clock battery is near end of life.

The battery needs to be replaced. This is not a field service

item. Contact customer service.

• Bad msg sent to display: reset PTE-2, contact customer service

Percent Error Operation messages

• % Err: in & out match: refer to section
• % Err: bad input range: refer to section
• % Err: bad output range:
• % Err: ext input of: press V/mA key to turn external channel on
• % Err: CH1 empty: no module is installed in Channel 1 bay
• % Err: CH2 empty: no module is installed in Channel 2 bay

SECTION 16 b.

TROUBLE SHOOTING CONT.